Remember the last time when your tubes showed condensation droplets after temperature incubation in a thermoblock or thermomixer? Eventually, you even had a performance loss in your enzymatic reaction.
What is condensation?
Condensation is defined as the change from the gas phase into the liquid phase. Condensation is the reverse process of vaporization. The condensation process occurs when the vapor is saturated and gets in contact with a cold (or colder) surface compared to its origin.
The higher the temperature delta between the warmest and the coldest spot within a closed system, the faster you will get condensation. The microtube is a perfectly closed system.
Many reactions need specific temperatures above room temperatures like 37°C, 42°C, 95°C, etc. As a result, an incubation step of 30 min at e.g. 65°C will end up with a layer of condensate droplets within the lid of your tube. You may think, one spin in the centrifuge and 100% of the sample are back at the bottom of the microtube. This process indeed helps to bring back the condensation droplets to your sample.
But there might still be an issue with the performed reaction which is done when you start your centrifugation step. Reagents need constant conditions for molecular processes like ligation, amplification, translation, restriction or whatever enzymatic reaction needs to be performed.
The reactivity of an enzyme depends on temperature, the availability of reaction partners but also on the salt concentration within the sample. Assuming a 25 µL enzymatic assay which has a salt concentration of e.g. 5 %. As the condensation process is limited to liquids, the calculated 3 µL condensate within the lid increases the salt concentration within the remaining sample at the bottom of the tube during the enzymatic reaction step by nearly 14 %.
(Rule of thumb: 5 % of 25 µL reaction volume: 1.25 µg salt, 22 µL of liquid incl. the salt remains after condensation, this is now 5.682 % of the total volume; resulting in 13.6 % higher salt concentration)
This increased salt concentration will probably not stop your reaction but might influence the efficiency of the enzyme, resulting in a different yield than possible under optimized conditions. Bottom line is, you are missing constant/ reproducible and efficient conditions.
There is a simple and convenient way to avoid condensation: Keep your sample liquids being the “coldest spot” in the reaction tube. For sure, the reaction temperature is set to e.g. 70°C. How to proceed?
One solution can be a counter-heating from the top of the vessel by a heated lid.